Online Orthophosphate Analyzer KOR600

Technical Specification of Orthophosphate Analyzer

 

ModelKOR600
Monitoring ParameterOrthophosphate
MethodAmmonium molybdate Spectrophotometer
Range(0.05-400) mg/L expandable
Display unit12232 dot matrix display, 7-inch LCD touch screen dual display
Measurement modeManual  measurement,     continuous    measurement,            periodic measurement,      fixed-point     measurement       and automatic

calibration measurement

Measurement periodAnywhere from 30 minutes to 24 hours
Zero Drift≤ ± 5% F. S
Range Drift≤ ± 5% F. S
Repeatability≤5%
Indication error≤ ± 5%
MBTF> 720 hours
Stored dataFive years of data storage
Communication interface2 Route 4-20mA,1 Route RS232,1 Route RS485
Ambient Temperature5-40 °C
Ambient Humidity(65 ± 20) % Rh
Power RequirementsRated voltage (220 ± 22) VAC frequency (50 ± 1) % Hz
Contour Dimension505 * 395 * 671(length * width * height unit mm)

 

 

Analyzer Principle and Application of Orthophosphate

 

 

The yellow phosphomolybdic heteropoly acid was produced by the reaction of orthophosphate with ammonium molybdate in acid solution, and then reduced to phosphomolybdic blue by ascorbic acid.

The main application scenarios are the monitoring of enterprise rainwater and sewage, the monitoring of total copper content in municipal pipe network, pumping station, groundwater, river water, lake water and sea water.

 

 